DT-300 and DT-100 Multi Pass Test Stand
- Product Details
- Company Profile
The test bench can automatically acquire the main performance parameters.
Test items:
Filtration efficiency (filtration accuracy)
Impurity storage capacity (pollution
capacity, service life)
Flow pressure drop characteristics
Test object:
Fuel Filter
Oil filter
Hydraulic filter
Standards
ISO 19438-2007 Diesel fuel and petrol filters for internal combustion engines -- Filtration efficiency using particle counting and contaminant retention capacity
ISO 19438-2007 Methods of test for full-flow lubricating oil filters for internal combustion engines - Part 12: Filtration efficiency using particle counting, and contaminant retention capacity
ISO 16889-2008 Hydraulic fluid power -- Filters -- Multi-pass method for evaluating filtration performance of a filter element
Features
The test bench can automatically acquire the main performance parameters.
The test bench adopts the automatic control, which can display the main test technical parameters such as pressure, flow rate, temperature, filtration efficiency and flow pressure drop in real time during the experiment. The test report can be automatically generated after the test and printed according to the need.
The test bench can monitor the test data during the experiment, and make the alarm and prompt function. The design of the corresponding security measures, can guarantee the safe and stable operation of the test bench.
The test bench shell is made of aluminum alloy, the panel is made of stainless steel, the word adopts wax word, and the connecting pipe adopts seamless stainless steel tube.
Note: The above parameters can be designed and manufactured according to the different requirements of customers.
Application
New product development.
Comparison with foreign product performance data.
Product quality control.
Third party testing, supervision agency.
Technical Parameters
Items | Parameters | |
Test flow | Test flow | 1L / min - 300L / min test cycle system flow segment division |
Flow rate of small flow section | 1L / min - 6L / min Note: The oil filter material system borrows this flow section. | |
Flow rate of medium flow section | 6L / min - 40L / min | |
Flow rate of large flow section | 40L / min - 300L / min | |
Sewage injection flow | 50ml / min - 300 ml / min | |
Online dilution system dilution ratio | 1:5 ~10 | |
Test cycle, sewage mixing system temperature | 40℃±1℃ | |
Online Particle Counting System Particle Counter | Upstream and downstream each 16 channels |