Automatic trace moisture analyzer, Karl Fischer Titrator
Automatic trace moisture analyzer, Karl Fischer Titrator

Automatic trace moisture analyzer, Karl Fischer Titrator

Member Details

Changsha Friend Experimental Analysis Instrument Co., Ltd

Changsha,Hunan,China

Manufacturer,Scientific Researcher

  • Product Details
  • Company Profile

Automatic trace moisture Analyzer uses Karl Fischer coulometric titration method

FDT-1371 Automatic micro moisture analyzer (Karl Fischer method)

Scope

This instrument conforms to the test standards GB/T 7600, ASTM D1744, GB/T 11133 and ASTM D6304 design and manufacture, it uses the Karl Fischer Coulometric Titrator Principle for the determination of trace moisture in the material., and easily contaminated electrode for reagents and reagent samples can be equipped with water-insoluble measurement sample was measured, the entire work process equipment automatically controlled by the computer with the corresponding trace moisture.


Features

1,320 × 240 dot matrix large LCD display, illustrated, intuitive and friendly

2, in the form of a bar graph shows the measured electrode signal, a visual indication of the water content of the electrolyte

3, real-time velocity curve depicting electrolysis time, you can determine whether the failure of the electrolyte according to this curve

4, the use of electrolyte blank current compensation, electrolyte balance drift compensation to correct the measurement results

5, large electrolytic current, can quickly achieve balance, into test status

6, time-stamped history, up to store up to 100, down retention time of 10 years

7, thermal micro-printer, print more beautiful and more efficient, with a spooling function

9, the introduction of imported XECOM US data processing chip, fast and accurate data processing;


Technical Parameters

Measurement accuracy: 

Testing of water between 10ug ~ 1000ug, error is less than ± 3ug

Testing the water is greater than 1000ug, error is less than ± 0.3%

Resolution: 0.1ug

Electrolytic current: 0 ~ 400mA

Standby power consumption: 6W

Dimensions: 350 × 260 × 180mm