Horizontal and Vertical Burning Tester
- Product Details
- Company Profile
Horizontal and Vertical Burning Tester Brief DescriptionHorizontal and Vertical Burning Tester with Standards IEC 60695-11-3/4 and UL 94-2006
Horizontal and Vertical Burning Tester:
Key Specifications/Special Features:
- Features:
- Main reference of horizontal and vertical burning test are UL 94-2006, IEC 60695-11-4, IEC 60695-11-3, GB/T5169-2008 series of standards
- It uses Bunsen burner with specified size and gas source (methane/propane)
- According to certain flame height and angle to burning horizontal/vertical combustion test facilities several times to assess flammability and fire risk with fire, burning duration and length of burning
- Horizontal and vertical burning tester is mainly for flammability l V-0, V-1, V-2, HB, 5V with grade material and for grading assessment
- Specifications:
- Burner:
- Diameter: 9.5, ±0.3mm
- Single gas and diameter: 9.5, ±0.3mm (single and mixed gas of Bunsen burner)
- Test point: 0, 20 and 45° manual switch
- Flame height: from 20, ±2mm to 180, ±10mm adjustable
- Burning time: 0 to 999.9s, ±0.1s adjustable
- After flame time: 0 to 999.9s, ±0.1s, automatically record, manual pause
- After burning time: 0 to 999.9s, ±0.1s, automatically record, manual pause
- Combustion gases: 98% of methane or propane gas (general LPG can be used to replace)
- Flow pressure: meter and pressure gauge (gas and air)
- Temperature test range: 0 to 1,000°C
- Flame temperature requirements: from 100, ±5°C up to 700, ±3°C within 44, ±2s and 54, ±2s
- Theromocouple (D): 0.5mm 2 import armor (type K) dual temperature system (50W) and 5VA/B (500W) grade
- Test background: black
- Test process: automatic control independent ventilation
- References: UL 94-2006, IEC 60695-11-3, IEC 60695-11-4, GB/T5169.15, 16, 17, 22-200815
- Studio size: 0.75m3, advanced fire board chassis (can be customized)
- Dimensions (W x D x H): 1,100 x 700 x 1,300mm
- Burner: